HY01
The GNST-HY01 infrared spectrophotometer is used to determine the oil after the water sample is extracted with tetrachloroethylene under the condition of pH ≤ 2; the petroleum is determined after the extract is adsorbed with magnesium silicate to remove polar substances such as animal and plant oils. The content of oil and petroleum is calculated based on the correction coefficient by the absorbance A2930, A2960 and A3030 at the wave numbers of 2930cm-1 (stretching vibration of C-H bond in CH2 group), 2960cm-1 (stretching vibration of C-H bond in CH3 group) and 3030cm-1 (stretching vibration of C-H bond in aromatic ring); the content of vegetable oil is the difference between the content of oil and petroleum.

• The whole operation is odorless, friendly to operators and the environment, and pollution-free.
• Fast detection speed, short stabilization time, and it only takes 10 seconds to detect a sample.
• Low reagent usage: solves the problems of large amounts of extractants and samples, strong odor, and complex purification and cleaning.
• It can be used on-site and in the laboratory, and is in an aluminum alloy box that can be carried to the site for testing.
• Low detection limit and high accuracy.

|
Name |
Parameters |
|
Detection method |
According to HJ637-2018 standard, "Infrared spectrophotometry" |
|
Method detection limit |
<0.04mg/L (sample volume 50mL, extract volume 10mL), <0.02mg/L (sample volume 500mL, extract volume 50mL), <0.002mg/L (sample volume 1000mL, extract volume 10mL) |
|
Instrument detection limit |
<0.10mg/L (3 times standard deviation of 11 times of tetrachloroethylene blank solution) |
|
Accuracy |
<±2% (20-100mg/L oil standard sample measurement 11 times) |
|
Repeatability |
RSD <0.2% (10-100mg/L oil standard sample measurement 11 times) RSD <0.6% (1-10mg/L oil standard sample measurement 11 times) |
|
Linear correlation coefficient |
No need to manually adjust the zero point, the software intelligently controls the adaptive automatic zeroing, no baseline drift |
|
Oil measurement scanning speed |
Full spectrum scan 10 seconds, three wave number scan 4 seconds |
|
Operating system |
2.0 version intelligent real-time operating system, 8-inch color touch screen voice prompt |
|
Data storage |
The instrument can store 262,000 records of data with spectrum, and there is no upper limit for data storage when connected to a computer |
|
Using environment |
Temperature range 5°℃ -45℃℃, humidity ≤90% |
|
Power supply mode |
Built-in 12V lithium battery, can be used for more than 24 hours after charging |
|
Host size |
Host size 37cm (length) x27cm (width) x 12cm (height) |
|
Host weight |
6.5kg |

Keywords:
HY01 Oil in water Laboratory
HY01 Oil in water Laboratory
Contact Us