Specifications
| TOC sensor | |
| Measuring range | 0.3...150mg/L |
| Principle | Two light sources,UV254 |
| Temperature range | 0-45℃ |
| Pressure range | 0...1bar |
| Process connection | NPT3/4 |
| Protection grade | Ip68 |
| Shell material | 316 Stainless steel |
| Installation | Immersion type |
| Cable length | 5m (standard) |
| Controller | |
| Measuring range | According to the sensor |
| Display | Back light, dot matrix LCD display |
| Resolution | 0.01/0.1 |
| Accuracy | ±5%FS |
| Repeatability | ±1.0%FS |
| Ambient temperature | 0~70℃ |
| Relative humidity | (65±20)% |
| Alarm output | double relay |
| Digital output | RS485 |
| Power supply | AC220V±10% or DC24V |
| Certification | CE |
Sensor installation

Sensor calibration
The sensor has been calibrated before shipment.
It is recommended to use the laboratory comparison method for calibration when calibrating.
Users are recommended to calibrate the sensor every 1 to 2 months.
Installation site



CE certification

Keywords:
Online TOC Meter
Related Downloads
Online TOC Meter
GENESITE TOC Meter in the range of 0.3...150mg/L are measured and displayed using a controller with RS485 Modbus RTU protocol .TOC sensor with self-cleaning functions.
Contact Us
Applications
Metallurgy, electric power, pharmaceuticals, chemical industry, petroleum, water treatment, food industry, etc.
Related Products
The GNST-24X dual temperature controlled digestion instrument has two temperature zones
Multi channel clamp with small mouth, sharp edges, and tightly connected skin tube.
The wavelength range of GNST-990UV visible spectrophotometer is 190-1100 nm
GNST-800A is a professional outdoor portable water quality detector with an integrated design.