Specifications
| Ammonia nitrogen sensor | |
| Measuring range | 0.1...18000ppm |
| Principle | Ion-selective method |
| Resolution | 0.01ppm |
| Accuracy | ±5%FS |
| Communication Protocol | RS485 Modbus RTU |
| Temperature range | 0-50℃ |
| Pressure range | 0...1bar |
| Process connection | 3/4,M39*1.5,G1 |
| Protection grade | Ip68 |
| Shell material | PPS,ABS,PC,316L SS |
| Installation | Immersion type/Flow through type |
| Cable length | 5m (standard) |
| Interference | Na +,K + |
| Certification | CE |
Sensor installation

Sensor calibration
It is recommended that users be put into use after calibration.
It is recommended to use the two-point method for calibration, usually with 1.000ppm solution calibrate the zero point, then the slopeis determined with the solution of 10.000ppm.
The electrode should be calibrated in fresh solution. The standard solution can be set, and the second standard solution should be larger than the solution to be tested.
Users are recommended to calibrate the sensor every 1 to 2 months.
Installation site



CE certification

Keywords:
Ammonia Nitrogen Sensor
Related Downloads
Ammonia Nitrogen Sensor
GENESITE Ammonia nitrogen Sensor in the range of 0.1...18000ppm with RS485 Modbus RTU protocol .The sensor with ATC functions .
Contact Us
Applications
Metallurgy, electric power, pharmaceuticals, chemical industry, petroleum, water treatment, food industry, etc.
Related Products
GENESITE Nitrate Sensor in the range of 0.1...14000ppm with RS485 Modbus RTU protocol .Nitrate sensor with ATC functions .
GENESITE Oil in Water Sensor in the range of 0...50ppm with RS485 Modbus RTU protocol .The sensor with self-cleaning functions .
GENESITE Chlorophyll Sensor in the range of 0...500ug/L with RS485 Modbus RTU protocol .Chlorphyll sensor with self-cleaning functions .
GENESITE Chroma Sensor in the range of 0...500 PCU with RS485 Modbus RTU protocol .The sensor with self-cleaning functions .