Specifications
| Nitrite sensor | |
| Measuring range | 0.1...14000ppm |
| Principle | Ion-selective method |
| Resolution | 0.01ppm |
| Accuracy | ±5%FS |
| Communication Protocol | RS485 Modbus RTU |
| Temperature range | 0-50℃ |
| Pressure range | 0...1bar |
| Process connection | 3/4,M39*1.5,G1 |
| Protection grade | Ip68 |
| Shell material | PPS,ABS,PC,316L SS |
| Installation | Immersion type |
| Cable length | 5m (standard) |
| Interference | ClO4−,I−,ClO2−,F− |
| Certification | CE |
Sensor installation

Sensor calibration
It is recommended that users be put into use after calibration.
It is recommended to use the two-point method for calibration, usually with 1.000ppm solution calibrate the zero point, then the slopeis determined with the solution of 10.000ppm.
The electrode should be calibrated in fresh solution. The standard solution can be set, and the second standard solution should be larger than the solution to be tested.
Users are recommended to calibrate the sensor every 1 to 2 months.
Installation site



CE certification

Keywords:
Nitrite Sensor
Related Downloads
Nitrite Sensor
GENESITE Nitrite Sensor in the range of 0.1...14000ppm with RS485 Modbus RTU protocol .The sensor with ATC functions .
Contact Us
Applications
Metallurgy, electric power, pharmaceuticals, chemical industry, petroleum, water treatment, food industry, etc.
Related Products
GENESITE TOC Sensor in the range of 0.3...150mg/L with RS485 Modbus RTU protocol .The sensor with self-cleaning functions .
GENESITE Ion Sensor contains the following ions: NH4+, NO3-, F-, Cl-, Ca2+, Cu2+, Na+, K+, Pb2+, I- etc. The Sensor with RS485 Modbus RTU protocol .
Electromagnetic flowmeter(Insertion-type )
GENESITE Insertion Type Electromagnetic flowmeter(GNST-YT-CRLLJ) with an ultra-compact transmitter.Nominal diameter: max. DN 3000 (120"). It has the characteristics of strong anti-interference ability and stable signal transmission.
Electromagnetic flowmeter(Split-type)
GENESITE Split-type Electromagnetic flowmeter(GNST-FTLL) with PTFE lining and rubber lining for you to choose from.Nominal diameter: DN 15 to 3000 (½ to 120").It can help you measure instantaneous flow.